20 Sure Temp® Plus Service Manual
4.1.3 Patient Temperature Out of Range
Patient Temperature Out of Range errors occur when the patient
temperature is above 110.0° F or below 80.0° F. During this error, the
entire body (head, torso, and legs) displays by blinking with a solid up or
down arrow and a thermometer bulb.
4.1.4 Probe Errors
The probe or probe connector can generate errors. Probe errors are
displayed as a probe with an X through it and an error code in the
temperature field. In most cases this will represent a broken probe that
needs replacing and the code will tell what error was detected.
Temperature measuring is prohibited until the error is cleared.
4.1.5 Instrument Errors
Instrument Circuitry errors are generated from internal test failures.
Instrument errors are displayed as an instrument with an X through it and an
error code in the temperature field. In most cases this will represent a broken
instrument that needs replacing and the code will tell what error was
detected. Error code numbers are available in Biotech mode under Error Log
(see page 11).

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn SureTemp Plus 692 Series and is the answer not in the manual?

Welch Allyn SureTemp Plus 692 Series Specifications

General IconGeneral
BrandWelch Allyn
ModelSureTemp Plus 692 Series
CategoryThermometer
LanguageEnglish

Related product manuals