EasyManua.ls Logo

Wheaton UniSpense PRO User Manual

Wheaton UniSpense PRO
29 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
WHEATON 800.225.1437 (U.S. & Canada Only) | 856.825.1100 | www.wheatonsci.com| P/N 50099073 Rev. 2/2/10 Lit. No.8734
Catalog Numbers:
W375040-A and W375040-B (100-120V)
and W375040-C through W375040-J (220-240V)
WHEATON | UniSpense
®
PRO
Peristaltic Pump
Instruction Manual
Instruction ManualInstruction Manual
Instruction Manual
Question and Answer IconNeed help?

Do you have a question about the Wheaton UniSpense PRO and is the answer not in the manual?

Wheaton UniSpense PRO Specifications

General IconGeneral
BrandWheaton
ModelUniSpense PRO
CategoryWater Pump
LanguageEnglish

Summary

Safety Symbols and General Safety

General Safety Rules and Precautions

Electrical Safety and Grounding

Focuses on crucial electrical safety, proper grounding, and avoiding adapters.

Specific Safety Considerations

Grounding and Electrical Connection Safety

Addresses grounding requirements, plug/outlet compatibility, and the dangers of improper connections.

Specifications – UniSpense® PRO

Initial Inspection and Unit Components

Unit Features and Controls

Identifies main components like the keypad, screen, I/O interface, and pump head.

Unit Control and User Interface

Installation and Initial Startup

Initial Startup Precautions

Advises on safe practices during initial startup, especially regarding rotating parts.

Pump Head and Tubing Installation

Keypad Functions and Interface Icons

Unit Setup and Configuration

Performing Calibration

Tubing Priming and Calibration Dispense

Completing Calibration Dispense

Covers performing a dispense, measuring volume, and entering it for calibration.

Batch and Single Dose Dispensing Setup

Executing and Adjusting Dispenses

Memory Storage and Recall

Storing and Recalling Parameters

Explains how to save and retrieve dispensing setups using memory registers.

Storing and Recalling Dispense Parameters

Saving Settings to Memory

Guides on selecting a register and storing current pump parameters.

Retrieving Saved Settings

Details how to load previously saved parameters and preview register contents.

Maintenance and Troubleshooting

Troubleshooting Safety Precautions

Emphasizes safety warnings for personnel performing repairs and maintenance.

Troubleshooting: Unit Operation Issues

Unit Not Operating or Motor Not Rotating

Addresses issues like blown fuses, low voltage, power switch, and motor control problems.

Pump Head and Dispensing Problems

Covers pump head engagement, tubing installation, and automatic cycling failures.

Motor Speed Inconsistencies

Addresses issues with the motor failing to maintain consistent rotation speed.

Troubleshooting: Dispensing Issues

Persistent Tubing Drip

Addresses causes of dripping after pump stop, such as worn or mismatched tubing.

Inaccurate Volume Delivery Causes

Covers issues like worn tubing, improper calibration, and clamp adjustments.

Fuse Replacement

Fuse Replacement Steps

Guides on safely disconnecting power, removing the fuse drawer, and installing new fuses.