20
14.3 Fuse Replacement
DANGER! BE CERTAIN THE UNIT IS DISCONNECTED FROM THE AC POWER
SOURCE.
1. Disconnect the cord-set from the AC power source.
2. Remove the fuse drawer by using a small flat screwdriver to lift
the tab.
3. Replace with new fuses:
108 - 120 vac unit use (2) 5x20mm, 250V, 2.0 AT fuses, be sure the correct cord set is used.
220-240 vac unit use (2) 5x20 mm 240V 1.0 AT fuses, be sure the correct cord set is used.