12 WIKA operating instructions model CPG1000
11568047.02 08/2012 GB/D
GB
3. Specications
3. Specications
Sensor technology
Measuring range mbar 0 … 70
1)
Overpressure limit mbar 200
Burst pressure mbar 350
Resolution mbar 0.001
Measuring range bar 0 … 1 0 ... 2 -1 ... +1
1)
-1 ... +2
1)
0 ... 7
Overpressure limit bar 4 4 4 4 14
Burst pressure bar 6 6 6 6 70
Resolution bar 0.0001
Measuring range bar 0 ... 20 0 ... 35 0 ... 70 0 ... 140 0 ... 200 0 ... 350 0 ... 700
Overpressure limit bar 42 70 140 210 420 700 1,100
Burst pressure bar 140 140 700 700 700 700 1,100
Resolution bar 0.001 0.01
Measuring range bar abs. 0 ... 1 0 ... 2 0 ... 7 0 ... 20
Overpressure limit bar abs. 4 4 14 42
Burst pressure bar abs. 35 35 70 140
Resolution bar abs. 0.0001 0.001
Accuracy 0.05 % FS
(0.1 % FS for measuring ranges 0 ... 70 mbar, -1 ... +1 bar and -1 ... +2 bar)
(0.25 % FS for vacuum measuring ranges with instruments from 7 bar to a max.
35 bar)
2)
Compensated temperature
range
0 ... 50 °C
Temperature coecients 0.005 % of the span/°C outside of 0 ... 50 °C
Type of pressure relative pressure, absolute pressure (up to 20 bar abs) and vacuum measuring
ranges
Pressure connection ¼ NPT male (incl. ¼ NPT female to G ½ male adapter)
3)
Pressure medium all liquids and gases compatible with 316 stainless steel
1)
1) Only use the measuring ranges marked in this way with clean non-corrosive gases.
2) For measuring instruments with a maximum measuring range of up to 2 bar, operation in the vacuum measuring range is limited
to -350 mbar. (Measuring ranges -1 ... +1 bar and -1 ... +2 bar are excluded.) Measuring ranges ≥ 7 bar can be loaded with a full
vacuum.
3) Adapter not included in delivery for North America.