40 WIKA operating instructions, model CPH6200
EN
11221780.03 07/2022 EN/DE
Reference pressure sensor model CPT6200
Gauge pressure bar -1 ... 0
1)
-1 ... 1.5
1)
-1 ... 3
1)
-1 ... 5
1)
-1 ... 9
1)
-1 ... 15
1)
-1 ... 24
1)
-1 ... 39
1)
0 ... 1 0 ... 1.6 0 ... 2.5 0 ... 4
0 ... 6 0 ... 10 0 ... 16 0 ... 25
0 ... 40 0 ... 60 0 ... 70 0 ... 100
0 ... 160 0 ... 250 0 ... 400 0 ... 600
0 ... 1,000
psi 0 ... 5 0 ... 10 0 ... 15 0 ... 20
0 ... 30 0 ... 50 0 ... 100 0 ... 150
0 ... 200 0 ... 300 0 ... 500 0 ... 1,000
0 ... 1,500 0 ... 2,000 0 ... 3,000 0 ... 6,000
0 ... 8,000 0 ... 14,500
Absolute pressure mbar
abs.
0 ... 250 0 ... 400 0 ... 600
bar
abs.
0 ... 1 0 ... 1.2 0 ... 1.6 0 ... 2.5
0 ... 4 0 ... 6 0 ... 10 0 ... 16
0 ... 25 0.8 ... 1.2
psi
abs.
0 ... 5 0 ... 10 0... 15 0 ... 20
0 ... 30 0 ... 50 0 ... 100 0 ... 150
0 ... 200
Overpressure safety 3 times; ≤ 25 bar
2 times; > 25 bar ... ≤ 600 bar
1.5 times; > 600 bar
3 times; ≤ 360 psi
2 times; > 360 psi ... ≤ 8,700 psi
1.5 times; > 8,700 psi
Process connection
G ½ B For all measuring ranges
G ½ B flush
3)
For measuring ranges > 1.6 ... < 1,000 bar and bar abs.
For measuring ranges > 20 ... < 14,500 psi and psi abs.
G 1 B flush
3)
For measuring ranges ≥ 0.1 ... ≤ 1.6 bar and bar abs.
For measuring ranges > 5 ... ≤ 20 psi and psi abs.
Adapter Various connection adapters on request
Material
Wetted parts Measuring ranges ≥ 0.1 ... 25 bar [≥ 1.45 ... 360 psi]
■
Stainless steel or
■
Elgiloy
®
Measuring ranges > 25 bar [> 360 psi]
■
Stainless steel and sealing from NBR or
■
Elgiloy
®
and sealing from NBR
10. Specifications