41WIKA operating instructions, model CPH6200
EN
11221780.03 07/2022 EN/DE
Reference pressure sensor model CPT6200
Wetted parts Measuring ranges < 100 mbar [< 1.45 psi]
Stainless steel
Silicon
Aluminium
Gold
Silicone
Oxygen version, measuring ranges ≥ 0.25 bar [≥ 0.4 psi]
Stainless steel or
Elgiloy
®
Flush version
Stainless steel with O-ring from NBR or
Stainless steel with O-ring from EPDM or
Hastelloy C4 with O-ring from NBR or
Hastelloy C4 with O-ring from EPDM
Pressure
transmission
medium
For measuring ranges to ≤ 16 bar [≤ 250 psi] synthetic oil
For flush version synthetic oil
For oxygen version halocarbon oil
Permissible ambient conditions
Medium
temperature
-30 ... +100 °C [-22 ... +212 °F]
-10 … +50 °C [14 ... 122 °F] (only for oxygen version)
Operating
temperature
-20 ... +80 °C [-4 ... +176 °F]
Storage temperature -40 ... +100 °C [-40 ... +212 °F]
Relative humidity 0 ... 95 % r. h. (non-condensing)
Case
Material Stainless steel
Ingress protection
IP65
IP67 when connected
Dimensions See technical drawing
Weight Approx. 220 g [0.49 lbs]
1) Not available as oxygen version.
2) Exclusively suitable for use with dry, gaseous and non-aggressive media. Not possible as flush version.
3) As an oxygen version or oil- and grease-free version, a flush version is not available.
10. Specifications