15.0 Serial Data Output :
The InfraRan analyzer outputs a constant asci stream on the rs232 and USB ports
when the RUN is initiated via the ANALYZE softkey menu.
The stream can be used to input specialized or custom applications on a host
PC to perform analysis functions external to the InfraRan analyzer. Or to perform
external logging functions.
The data stream :
13MAR06, 12:29:11, 2.7553, 0, 0.0077, 2.7535, 20.67, 123
Field 1 2 3 4 5 6 7 8
1. Date, DDMMMYY.
2. Time, HH:MM:SS.
3. Buffered voltage value from detector.
4. Value displayed on InfraRan VFD display.
5. Absorbance value.
6. Unbuffered voltage value from detector.
7. Temperature value from internal thermistor ( Celsius ).
8. Analyzer I.D. ( this field is only enabled when a value is entered via the
SETUP - CALIBRATE menu )