Wilks Enterprise Inc InfraRan Instruction Manual
31
3. The InfraRan is Unresponsive to Anesthetics
Remove the sample wand and filters. Cover first the inlet and then the outlet fitting with
a finger. Verify that you can feel a slight vacuum at the inlet and a slight positive
pressure at the outlet. If there is no indication of either pressure or vacuum, check that
the InfraRan is in Run mode. Press Control->Pump. If Pump Off is highlighted press
the key under Pump On to turn the pump on. You should be able to hear the pump and
feel vacuum at the inlet and pressure at the outlet.
If the pump fails to operate or if there is either no vacuum at the inlet or pressure at the
outlet, the InfraRan must be returned to Wilks Enterprise for repair.
4. Display is Blank or Garbled
If the display is not functioning properly the InfraRan must be returned to Wilks
Enterprise for repair.
5. Some or All Keypad Keys Do Not Function
The InfraRan must be returned to Wilks Enterprise for repair.