4. Setting Up The Bed
• Make sure that the fl oor surface is suitable, so that no damage can occur to the fl oor. Floors that
are, for example, too soft, unsealed or with imperfections are not suitable. These fl oors are often not
suitable: soft wooden fl oors, porous and soft stone fl oors, carpets with foam backing, soft linoleum or
similar fl oor covering. If in doubt, please contact your fl ooring company and the Technical Customer
Service of Wissner-Bosserhoff GmbH.
• Ensure that the bed stands freely in the room so that there is a suffi cient safety clearance all
around the bed for all adjustment positions (including the tilting function). For the extreme
positions of the bed, an additional safety clearance of 2.5cm from objects, walls and installations
must be taken into account, as protection from entrapment. The safety clearance of the head end
from objects, walls and installations is at least 20cm.
• Be aware of the existence of any mechanical and electrotechnical protective measures, such as
wall-mounted bumpers and corner guards, RCB switches etc.
• Pay attention to the whereabouts of useful wall sockets for the mains connection for the bed.
Positions are unsuitable that could lead to collisions or stretching when the bed is adjusted, or in
which the mains cable stretches too far across the fl oor or even under the bed.
• In general, the use of extension cables or multiple adapters that may lie unsecured on the fl oor
of the room should be avoided.
• Magnetic objects or magnetic fi elds must be kept at a distance
5. Technical Data For The Bed / Nameplate1
For an explanation of the respective nameplate details – see Appendix B
• Outer dimensions: approx. 208 x 106 cm (depending on the wooden linning)
• Nominal dimension of
patient surface : 90 x 200 cm
• Height adjustment: 36 to 80 cm depending on the model (measured central
section)
• Backrest adjustment: 70° / electric sliding backrest SBR 11.5 cm
• Thigh section adjustment: 32° / electrical sliding thigh rest 4 cm
• Safe workload LF: Up to 225 kg (175kg patient +30kg mattress + 20 kg linen
and accessories) depending on model and fi ttings
• Tilting: approx. sitting position 13° /legs up position 15°
• Bed weight: approx. 125 kg depending on model and fi ttings
• Protection type: IPX4
• Noise level: approx 57 dB/A
• Anwendungsumgebung:
• the environment temperature lies in the range from + 10°C to + 40°C
• the relative humidity lies in the range from 30% to 75%
• the atmospheric pressure lies in the range from 700 hPa to 1060 hPa