Wissner-Bosserhoff Carisma Bed, Instruction & Safety Manual
T 08 9351 7500 E info@unicarehealth.com.au W www.unicarehealth.com.au
5.1. Electromotor Patient Surface
• Electrical connection: 230 V 1,25 A 50 Hz
• Protection type IPX4
• Class of protection II
• Output voltage 24 V / 32 V depending on model variant
(please read nameplate)
• Operating time of motors max. 2 min. adjustment operation, 18 min. pause
• Power consumption up to approx. 320VA peak load
• Adjustment speed approx. 5.4 mm/sec.
2
• Pressure force before redirection max. 3500N
1
All technical data are nominal data. Construction and manufacturing tolerances are in accordance
with EN ISO 14253-1 and GUM-1:95
2
Approximate values according to the model design and load value
5.2. Electric Motor Height Adjustment
• Max. voltage in idling mode 24 V
Current limiting (ÜSA) max.10 A
• Max. lifting force, direct 2x6000 N
• Max. tensile load 3,000N
• Adjustment speedv 10 – 14 mm/s
(depending on model fi ttings and load)
• Protection type
2
IPX4
• Class of protection III protective extra-low voltage
• Operating time of motors max. 2 min. adjustment operation, 18 min. pause