Volume mode operation
Infusion mode selection
Press in stop, display infusion mode interface. If the present mode is volume mode, so not need to
select mode.
Choose “volume mode with up or down key.
Press
, then return to infusion interface after volume mode is chosen.
VTBI/Rate/Time
Move the cursor to corresponding input boxes of rate, VTBI and time with the up or down key.
Set rate.
Set VTBI
Set time. (if need)
Press
Notes: set every two of these three parameters, the rest one will be calculated automatically, the machine would
calculate another parameter with the two last input parameters. And infusion would begin according to the
two last input parameters.