Chapter 9 Routine check
In routine use, please perform routine maintenance and check according to the following
table:
Check whether the housing of WIT-601 series infusion pump is intact, whether
cracks, broken and deformed exist.
Boot and do not connect AC, the icon“ ” flashes and accompanies light-sound
alarm, by this moment connect AC, if the icon appears, the AC supply is
normal, if not, the AC supply is abnormal.
Finger-type
peristaltic pump
Correctly install infusion pump and clip infusion set, then boot.
Press “start” to start infusion. Check whether liquid drops or not, and check
whether the finger-type peristaltic pump turns abnormally.
After booting, have not installed IV Set, but “bubble” alarms.
Install IV Set with liquid, close door, press “clear”, “bubble” alarm flash should
Correctly install infusion pump and clip infusion set, then boot.
Open pump door, door-open alarms, and no liquid drops.
Close pump door and press “clear” key, door-open alarm disappears, and no
Correctly install infusion pump and clip infusion set, then boot.
Start infusion and observe when every liquid drop drops, drop sensor would flash
Take out drop sensor in process of infusion, stop infusion, prompting “drop
Correctly install infusion pump and clip infusion set, then boot.
Set infusion rate 100ml/h, occlude manually the bottom infusion set after 5
minutes, and then stop infusion after about 1 minute, “down occlusion”
sound-light alarms.
Correctly install infusion pump and clip infusion set, then boot, purge the bubbles
Set rate to 90ml/h, VTBI to 15ml, start infusion, measure the liquid
volume flowed out from machine with a measuring cup.
Check whether the liquid volume flowed out is within 15ml±5%.
note
:
Carry out the above checks in specified using phase or check stage, if find error, please stop use and contact supplier as soon
as possible.