Clear the VI
Move the cursor to the VI box, press . The VI is reset.
Battery life check
In process of infusion, can check the runtime of battery under present infusion rate.
Choose “battery life”, press .
Notes
:
Because the runtime of battery depends on the working environment and the aging degree of
battery, there may be a deviation between the runtime checked and the real runtime, so medical staff
should pay attention at any time to the working status of machine when in use without AC
connection. In case of near empty of battery, should dispose in time.
Setting fluid warmer
Press in progress of infusion.
Choose the menu of fluid warmer, press .
Choose the menu of heating range, press .
Press to return to infusion interface.
Pre-alarm setting of infusion completion
Choose the menu of infusion completion pre-alarm, press .
Input the time need be advanced(5~30mins or input 0 to turn off pre-alarm), press .
Press to return to infusion interface.
Bottom occlusion pre-alarm setup
Choose the menu of bottom pressure pre-alarm, press .
Choose “ON” or “OFF” pressure pre-alarm, press .
Press to return to infusion interface.
key lock (auto)
Choose the menu of system parameters, press .
Choose the menu of keypad lock setting, press .
Input the time of automatic keypad lock (input 0 to cancel automatic keypad lock), press .
Press to return to infusion interface.
key lock (manual)
Long press key.then the keypad will lock,