press key again.then the keypad will unlock,
System parameters setup
System parameters in pause status
Choose “system parameters”, press .
Sensitivity setting of up
occlusion detection
3 level options: high, median and low
When not easy to prompt occlusion alarm but it please
When easy to occlusion alarm but it prompts in a long
time or it doesn’t prompt, please set to a lower level.
Sensitivity setting of bottom
occlusion detection
20 level options: 200-1200mmHg
When not easy to occlusion alarm but it prompts,
please set to a higher level.
When easy to occlusion alarm but it prompts in a long
time or it doesn’t prompt, please set to a lower level.
Sensitivity setting of bubble
detection
When big bubble appears but alarm doesn’t prompt,
please choose the level of small number.
When bubble alarm prompt in high frequency, please
choose the level of larger number.
Pre-alarm setting of infusion
completed
Pre-alarm time setting of infusion completed: 5~30min
Can set to no need pre-alarm
Time setting of pause alarm
Manual keypad lock setting
long press keypad lock to lock keypad after setting
Automatic keypad lock time
setting
Lock keypad automatically after 30s~1min settable
Choose suitable sound volume according to the environment