Set system time of machine
Can input year, month, date; hour, minute, second
Brightness adjustment of
LCD
Use up and down keys to adjust the brightness
System parameters check or setting in process of infusion
Sensitivity setting of up
occlusion detection
Sensitivity setting of up
occlusion detection
3 level options: high, median and low
When not easy to prompt occlusion alarm but
it prompts, lease set to a higher level.
When easy to prompt occlusion alarm but it
prompts in a long time or it doesn’t prompt,
please set to a lower level.
Sensitivity setting of
bottom occlusion detection
Sensitivity setting of
bottom occlusion detection
20 level options:200-1200mmHg
When not easy to prompt occlusion alarm but
it prompts, please set to a higher level.
When easy to prompt occlusion alarm but it
prompts in a long time or it doesn’t prompt,
please set to a lower level.
Sensitivity setting of bubble
detection
When big bubble appears but alarm doesn’t
prompt, please choose the level of small number.
When bubble alarm prompt in high
frequency, please choose the level of larger
number.
Pre-alarm setting of
infusion completed
Pre-alarm setting of
infusion completed
Pre-alarm time setting of infusion completed:
Can set to no need pre-alarm
Anti-bolus (Pressure
release)
The pressure of IV set
goes higher and the
pump will stop and the
motor will reverse to
reduce the pressure
Working with pressure sensor and control by the
pump software
Check present battery run time