Chapter 8 Security characteristic and alarm
Trouble shooting
Close the pump door and lock.
Infusion set is not properly
installed
Properly install the infusion set and make sure it is
mounted to the slot of bubble detector
Flick the infusion tube or press Bolus key to purge the
bubbles
Press the Clear key to cancel the alarm manually.
Replace new liquid, re-set the volume for
Infusion completed and shutdown.
Cancel pre-alarm if not needed
Liquid dripping is not
detected
Check whether the infusion bottle/bag is empty;
Check whether Murphy's dropper is misty or too
tilted, liquid flows down along the wall rather
than dripping
Infusion inlet is not open
Up pressure sensor is
interfered
Check for strong electromagnetic interference
around, re-start infusion pump after removal of
the source of interference
Up Pressure
Sensor Failure
Connection of up pressure
sensor looses
Contact the manufacturer for maintenance
Check whether the infusion tube is folded, and
Check whether the infusion tube/needle is
Infusion filter is blocked
Check whether for foreign objects are in filter, if
any, replace infusion set
Infusion set is too hard or
too thick
Replace infusion set or reduce blocking
Tube is blocked slightly,
the pressure going up
Check whether the tube is blocked
Set pressure level to a higher level