Can close pressure pre-alarm if not needed
Bottom pressure sensor is
interfered
Check for strong electromagnetic interference
around, re-start infusion pump after removal of
the source of interference
Bottom pressure sensor
fails
Pressure sensor is damaged, contact the
manufacturer for maintenance
Down Pressure
Sensor Failure
connection of bottom
pressure sensor looses
Contact the manufacturer for maintenance
IV set has not been
calibrated
Re-calibrate the IV set according to user’s manual
Tube type setting error
Not re-set the rate
parameters after replacing
the tube
Set correctly the tube type
Set correctly the parameters
Connection of drop sensor
is off or looses
Contact the manufacturer for maintenance
Sensor is not properly
clipped
Clip it properly on Murphy’s dropper according
Check the infusion line is installed correctly
Check whether the drip rate sensor is subjected to
strong interference/shaking
Check whether the door pressing plate is loose
Fluid warmer connection
looses
Contact the manufacturer for maintenance
Battery power is low and
cannot be charged
normally
Battery connection looses
Install the batteries correctly
Connect AC supply to charge the battery
Contact the manufacturer for maintenance
Stepper motor cannot run
normally
Infusion set is not properly installed, so that the
motor cannot drive the peristaltic pump normally.
Infusion pump fails, contact the manufacturer for
Infusion suspended over 2
min
Press "Clear" key to cancel the alarm and press
"Start" to continue the infusion
AC supply is not
connected correctly
Plug the AC supply correctly
Contact the manufacturer for maintenance
PS: if there is a single fault, the reason may be the infusion volume is less than 0.7m