Fault description Cause Remedy
The LED in the base of the
device does not light up/
flash (ChronoMaster Air
only)
The rechargeable bat-
teries are not charged.
Ensure that the rechargeable batteries
have been charged.
Connect the device to a power supply
using the USB cable.
Move the slide switch from “ON” to “OFF”
and back to “ON”.
The device is not listed in
the selection list for
channel assignment.
The device is not con-
nected to the PC or
tablet computer cor-
rectly.
Move the slide switch from “ON” to “OFF”
and back to “ON”.
Close the selection window for channel
assignment, and open it again.
Ensure that the device is connected to the
right PC or tablet computer.
The Bluetooth connection
is interrupted (Chrono-
Master Air only)
Unstable Bluetooth
connection.
Place the ChronoMaster Air closer to the
PC or tablet computer.
Use the external Bluetooth dongle “Laird”
(item no. JB15-BT851) for a more stable
connection.
The capacity of the
rechargeable batteries is
not displayed correctly
(ChronoMaster Air only)
The charge status dis-
play was not calibrated
after replacing the
rechargeable batteries.
Recalibrating the charge status display
(
Ä
“Recalibrating the charge status display”
on page 38).
24.03.2021 Precision measuring device ChronoMaster / ChronoMaster Air 41