IM 01E20D01-01E
5-10
5.  MAINTENANCE
5.9 Troubleshooting
Although magnetic flowmeters rarely require maintenance, failures may occur when the
instrument is not operated correctly. This section describes troubleshooting procedures
where the cause of the breakdown is identified through receiver indication.
5.9.1 No Indication
Is an error being displayed? Check the converter display.
NO
YES
NO
NO
YES
NO
YES
NO
Is an error being
displayed?
Is anything else
being displayed?
YES
YES
Check whether the flowtube 
is filled with fluid.
Is the flowtube
full of fluid?
Remove the tube and 
inspect the inside of the pipe.
Is foreign 
matter adhering to 
the walls?
Does the
arrow on the tube match the
flow direction?
Refer to the “Alarm Functions” in 
the AXFA11, AXFA14, and AXF 
Integral Flowmeter [Software 
Edition] user manuals.
Perform either of the following:
a. Reinstall the flowtube in 
order to make the arrow and 
flow direction match.
b. Reverse the setting for the 
flow direction.
Change the piping or the 
mounting condition so that 
the tube fills with the fluid.
Remove the foreign matter.
START
Contact a service center.
Check the resistance between 
the flowtube’s A, B, and C 
terminals using a circuit tester.
*
1
Is the
resistance between A-C or B-C 
100 M or more?
NO
YES
F0521.EPS
*1: Follow the warning in Section 5.7 
before checking. The integral 
flowmeter's resistance cannot be 
checked.