4-81
IM DL350-17EN
Commands
4
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
VDIV
Function Sets or queries the V/div setting of a 4-CH
module, 16-CH voltage input module, 16-CH
temperature/voltage input module.
Syntax
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
VDIV {<Voltage>}
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
VDIV?
4-CH module (720254)
<x1> = 1 to 4
<x2> = 1, 2
<Voltage> = 10 mV to 50 V
16-CH Voltage Input Module (720220)
<x1> = 1, 3
<x2> = 1 to 16
<Voltage> = 200 mV to 2 V
16-CH Temperature/Voltage Input Module
(720221)
<x1> = 1, 3
<x2> = 1 to 16
<Voltage> = 1 mV to 2 V
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
VGAin
Function Sets or queries the gain setting of a 4-CH
module.
Syntax
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
VGAin {<NRf>}
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
VGAin?
<x1> = 1 to 4
<x2> = 1 to 2
<NRf> = 0.9500 to 1.0500
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
ZOOM
Function Sets or queries the vertical zoom factor of a 4-CH
module, 16-CH voltage input module, 16-CH
temperature/voltage input module.
Syntax
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
ZOOM {<NRf>}
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
ZOOM?
4-CH module (720254)
<x1> = 1 to 4
<x2> = 1, 2
16-CH Voltage Input Module (720220)
<x1> = 1, 3
<x2> = 1 to 16
16-CH Temperature/Voltage Input Module
(720221)
<x1> = 1, 3
<x2> = 1 to 16
<NRf> = 0.1, 0.111, 0.125, 0.143, 0.167, 0.2, 0.25,
0.33, 0.4, 0.5, 0.556, 0.625, 0.667, 0.714,
0.8, 0.833, 1, 1.11, 1.25, 1.33, 1.43, 1.67,
2, 2.22, 2.5, 3.33, 4, 5, 6.67, 8, 10, 12.5,
16.7, 20, 25, 40, 50, 100
:CHANnel<x>[:VOLTage]:VARiable
Function Sets or queries how the vertical scale is set for a
module that can measure analog voltage.
Syntax
:CHANnel<x>[:VOLTage]:
VARiable {<Boolean>}
:CHANnel<x>[:VOLTage]:VARiable?
<x> = 1 to 4
Example
:CHANNEL1:VOLTAGE:VARIABLE 0
:CHANNEL1:VOLTAGE:VARIABLE?
-> :CHANNEL1:VOLTAGE:VARIABLE 0
:CHANnel<x>[:VOLTage]:VDIV
Function Sets or queries the V/div setting of a module that
can measure analog voltage.
Syntax
:CHANnel<x>[:VOLTage]:
VDIV {<Voltage>|<Current>}
:CHANnel<x>[:VOLTage]:VDIV?
<x> = 1 to 4
<Voltage> = See the table below.
Example
:CHANNEL1:VOLTAGE:VDIV 50
:CHANNEL1:VOLTAGE:VDIV?
-> :CHANNEL1:VOLTAGE:VDIV 50
Description Selectable Ranges by Voltage Measurement
Module Type (1:1 probe attenuation)
Type Selectable Range of VDIV
720250 5 mV to 20 V
720268 20 mV to 200 V
701261, 701262 5 mV to 20 V
701265 0.1 mV to 10 V
701275 5 mV to 10 V
720211 10 mV to 20 V
720266 0.1 mV to 20 V
:CHANnel<x>[:VOLTage]:VGAin
Function Sets or queries the gain adjustment of a module
that can measure analog voltage.
Syntax
:CHANnel<x>[:VOLTage]:VGAin {<NRf>}
:CHANnel<x>[:VOLTage]:VGAin?
<x> = 1 to 4
<NRf> = 0.9500 to 1.0500
Example
:CHANNEL1:VOLTAGE:VGAIN 1.01
:CHANNEL1:VOLTAGE:VGAIN?
-> :CHANNEL1:VOLTAGE:VGAIN 1.01
Description This command is valid when any of the following
modules is installed and the input coupling is set
to DC.
720211 (HS100M12), 720250 (HS10M12),
720254 (4CH 1M16)
4.6 CHANnel Group