4-41
IM DL350-17EN
Commands
4
:CHANnel<x>:CAN?
Function Queries all settings of a module that can monitor
CAN bus signals.
Syntax
:CHANnel<x>:CAN?
CAN Bus Monitor Module (720240) or
the CAN/CAN FD monitor module (720242)
<x> = 1 to 4
CAN & LIN Bus Monitor Module (720241)
<x> = 1, 3
:CHANnel<x>:CAN:OSOut?
Function Queries all one-shot output settings of the
specified port of a module that can monitor CAN
bus signals.
Syntax
:CHANnel<x>:CAN:OSOut?
CAN Bus Monitor Module (720240) or
the CAN/CAN FD monitor module (720242)
<x> = 1 to 4
CAN & LIN Bus Monitor Module (720241)
<x> = 1, 3
:CHANnel<x1>:CAN:OSOut:DATA<x2>
Function Sets or queries the one-shot output data of the
specified port of a module that can monitor CAN
bus signals.
Syntax
:CHANnel<x1>:CAN:OSOut:
DATA<x2> {<String>}
For the CAN bus monitor module (720240)
<x1> = 1 to 4, <x2> = 1 to 8
For the CAN/CAN FD monitor module (720242)
<x1> = 1 to 4, <x2> = 1 to 64
For the CAN & LIN monitor module (720241)
<x1> = 1, 3, <x2> = 1 to 8
<String> = “0” to “FF”
Example
:CHANNEL3:CAN:OSOUT:DATA1 "FF"
:CHANNEL3:CAN:OSOUT:DATA1?
-> :CHANNEL3:CAN:OSOUT:DATA1 "FF"
Description If a string outside the range is specified, error
151, “Invalid string data,” will occur.
:CHANnel<x>:CAN:OSOut:DLC (Data
Length Code)
Function Sets or queries the one-shot output data length
of the specified port of a module that can monitor
CAN bus signals.
Syntax
:CHANnel<x>:CAN:OSOut:DLC {<NRf>}
CAN Bus Monitor Module (720240) or
the CAN/CAN FD monitor module (720242)
<x> = 1 to 4
CAN & LIN Bus Monitor Module (720241)
<x> = 1, 3
<NRf> = 0 to 15
Example
:CHANNEL3:CAN:OSOUT:DLC 15
:CHANNEL3:CAN:OSOUT:DLC?
-> :CHANNEL3:CAN:OSOUT:DLC 15
:CHANnel<x>:ACCL:UNIT
Function Sets or queries the unit string when the coupling
of an acceleration/voltage module is set to
acceleration.
Syntax
:CHANnel<x>:ACCL:UNIT {<String>}
:CHANnel<x>:ACCL:UNIT?
<x> = 1 to 4
<String> = Up to 4 characters
Example
:CHANNEL1:ACCL:UNIT “m/s2”
:CHANNEL1:ACCL:UNIT?
-> :CHANNEL1:ACCL:UNIT “m/s2”
:CHANnel<x>:ACCL:VARiable
Function Sets or queries the scale setting mode when the
coupling of an acceleration/voltage module is set
to acceleration.
Syntax
:CHANnel<x>:ACCL:
VARiable {<Boolean>}
:CHANnel<x>:ACCL:VARiable?
<x> = 1 to 4
Example
:CHANNEL1:ACCL:VARIABLE 0
:CHANNEL1:ACCL:VARIABLE?
-> :CHANNEL1:ACCL:VARIABLE 0
:CHANnel<x>:ACCL:ZOOM
Function Sets or queries the vertical zoom factor when the
coupling of an acceleration/voltage module is set
to acceleration.
Syntax
:CHANnel<x>:ACCL:ZOOM {<NRf>}
:CHANnel<x>:ACCL:ZOOM?
<x> = 1 to 4
<NRf> = 0.5, 0.556, 0.625, 0.667, 0.714, 0.8,
0.833, 1, 1.11, 1.25, 1.33, 1.43, 1.67, 2,
2.22, 2.5, 3.33, 4, 5, 6.67, 8, 10, 12.5,
16.7, 20, 25, 40, 50
Example
:CHANNEL1:ACCL:ZOOM 1.000
:CHANNEL1:ACCL:ZOOM?
-> :CHANNEL1:ACCL:ZOOM 1.000
4.6 CHANnel Group