7.13 General Specifications
Item Specifications
Warm-up time Approx. 30 minutes
Operating environment Temperature: 5°C to 40°C
Humidity: 20%RH to 80%RH (no condensation)
Elevation: 2000 m or less
Installation location Indoors
Storage environment Temperature: -25°C to 60°C
Humidity: 20%RH to 80%RH (no condensation)
Rated supply voltage 100 VAC to 240 VAC
Permitted supply voltage range
90 VAC to 264 VAC
Rated supply frequency 50/60 Hz
Permitted supply voltage
frequency range
48 Hz to 63 Hz
Maximum power
consumption
WT310E, WT310EH: 50 VA, WT332E/WT333E: 70 VA
External dimensions
(excluding protrusions)
WT310E, WT310EH: Approx. 213 (W) × 88 (H) × 379 (D) mm
WT332E/WT333E: Approx. 213 (W) × 132 (H) × 379 (D) mm
Weight WT310E, WT310EH: Approx. 3 kg
WT332E/WT333E: Approx. 5 kg
Battery backup Setup parameters are backed up with a lithium battery.
Standard accessories Power cord: 1 piece (If the suffix code is -Y, a power cord is not included.)
D/A cable: 1 piece (included in products with the /DA4, /DA12 options)
Current input protection cover: 1 piece that matches the main unit
Hind feet rubber: 1 set
User’s manual: 1 piece (this manual)
Safety terminal adapter:
WT310E, WT310EH: Red/Black 1 set
WT332E: Red/Black 2 sets
WT333E: Red/Black 3 sets
Safety standard
1
Compliant standard: EN 61010-1, EN 61010-2-030
Installation category (overvoltage category) CAT II
2
Measurement Category CAT II
3
Pollution degree 2
4
Emission
1
Compliant standards
EN 61326-1 Class A
EN 55011 ClassA, Group1
EN 61000-3-2
EN 61000-3-3
EMC Regulatory Arrangement in Australia and New Zealand EN 55011 Class A, Group 1
Korea Electromagnetic Conformity Standard ( 한국 전자파적합성기준 )
This product is a Class A (for industrial environment) product. Operation of this product in a
residential area may cause radio interference in which case the user will be required to correct
the interference.
Cable conditions
Serial (RS-232) interface connector
Use shielded serial cables.
5
GP-IB interface connector
Use shielded GP-IB cables.
5
USB port (PC)
Use shielded USB cables.
5
Ethernet port
Use category 5 or better Ethernet cables (STP).
6
D/A output terminal
Use shielded cables.
5
Connecting cables may cause radio interference in which case the user is required to
correct the interference.