Static mode
This function is mainly for professional personnel
to enter the static mode to test the monitor
through standard pressure gauge.
Warning: Normal users don't need to know
this function and also do not operate. The
company will not take any responsibility for
damage caused by this operation.
Press the '' '' button after battery installation,
then the screen will show the '' '' icon, which
means the system is in zero testing. Several
seconds later, the '' '' icon disappears and the
air pump starts inflating at the same time, which
indicates the test ended. Then press the '' ''
button to stop inflating and take out the
batteries to enter the next step.
System restores
Note: It must restore the system before
entering the static mode, otherwise it may
cause inaccurate results.
Press the '' '' button and hold, meanwhile
install the batteries. Hold on for about 3 seconds
then release the '' '' button. Then screen will
show the pressure value '' '' . Now the system
has restored and entered the static mode. Now
can take the static test.
Entering the static mode
23
22
Static mode
After entering the static mode, if
the screen still doesn't show '' '',
please operate again as the
System restore. Please contact
with the local distributor if it still
does not work.
Tips!
Note: When the measured
pressure exceeds 300mmHg during
static test, pressure value and " "
are displayed, indicating the
measurement result exceeds the
rated range and may be inaccurate.
The static mode
The monitor will automatically power off if
there is no operation in 4 minutes.
Method of verifying calibration
The verification system is determined by
applying a '' Y '' adapter to the pressure line
and attaching a reference standard.
Enter into the static mode, read the device and
the reference gauge simultaneously, the error
of 3mmHg is normal by reducing the pressure
from 300mmHg to zero at a rate of
3mmHg/s±1mmHg/s.
If the error out of 3mmHg, please contact the
manufacturer for calibration.