Electromagnetic compatibility information
3332
Electromagnetic compatibility information
Portable
peripherals such as antenna cables and external
antennas) should be used no closer than 30cm (12
inches) to any part of the SPHYGMOMANOMETER,
including cables specified by the manufacturer.
Otherwise, degradation of the performance of this
equipment could result.
RF communications equipment (including
This equipment might not offer adequate
to radio-frequency communication services. The
user might need to take mitigation measures,
such as relocating or re-orienting the equipment.
protection
When the instrument is in use,never put it near
other instruments or stack it on other instrument.
if you have to put it near other instruments or
instruments, please inspect and verify if the
instrument could run normally.
Essential performance
1. Limits of the error of the manometer:
Over the temperature range of 5℃ to 40℃ and
the relative humidity range of 15% to 90% (non-
condensing), the maximum error for the
measurement of the CUFF pressure at any point of
the NOMINAL measurement range shall be less
than or equal to ±3mmHg (±0.4kPa) of the
reading.
2. Reproducibility of the BLOOD PRESSURE
DETERMINATION:
The laboratory reproducibility of the BLOOD
PRESSURE DETERMINATION of the AUTOMATED
SPHYGMOMANOMETER shall be less than or equal
to 3.0mmHg (0.4 kPa).
There is the potentia risk of radio frequency interference
between the device and other devices. If there is, please
find out the problems and take the following measures:
(1) Turn off the device, and turn on again.
(2) Change the direction of the device.
(3) Keep the product away from the interferential devices.
WARNING: The Operator should not use the
system and should inform the customer service, if
the ESSENTIAL PERFORMANCE is lost or degraded
due to EM DISTURBANCES.
WARNING: Use of accessories and cables other
than those specified or provided by the
manufacturer of this equipment could result in
increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and
result in improper operation.
List of cables and accessories is as follows:
Cable Name
DC power
supply cord
Cable Length
≤1.2m
Cable Shielded
Unshielded
Comments
None
Table 1 For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture s declaration- electromagnetic emission
'
Emission test
Compliance
Class BRF emissions CISPR 11
Class A
Complies
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/flicker
emissions IEC 61000-3-3
RF emissions CISPR 11
Group1
The YE660F Electronic Blood Pressure Monitor is intended for use in
the the electromagnetic environment specified below. The customer or
user should assure that of YE660F Electronic Blood Pressure Monitor
it is used in such an environment.