5. Blood pressure variations?
An individual's blood pressure varies greatly on a
daily and seasonal basis. It may vary by 30 to 50
mmHg due to various conditions during the day. In
hypertensive individuals variations are even more
pronounced. Normally, the blood pressure rises
while at work or play and falls to its lowest levels
during sleep. So do not be overly concerned by
the results of one measurement. Take
measurements at the same time every day using
the procedure described in this manual to get to
know your normal blood pressure. Regular
readings give a more comprehensive blood
pressure history. Be sure to note date and time
when recording your blood pressure. Consult
doctor to interpret the blood pressure data.
6. When is the best time to measure blood
pressure?
After urination, before breakfast in the morning.
Before sleeping at night.
Before taking medicine.
Please keep a stable body state and mind every
time measuring. We suggest taking
measurements at the same time every day.
Common fault and trouble shooting
Common fault
Solutions
Check the batteries are installed correctly
Replace new batteries
Multiple occurrences of
measuring failure, or
measured value is low
(or high)
Check the connection and winding of cuff
Check if the cuff winded too tight or
too loose. Take off your clothes if rolled
too tight
Please ensure a quiet, relaxed body
state. Deep breathing to relax yourself
before measurement
Please read the "blood pressure
variations" carefully
The monitor is in good
condition, but the each
measuring result are different
Pump works, but the pressure
doesn't rise
Check whether the cuff has connected
well
The value is different from
that measured at a clinic or
doctor's office
Write down the value every day,
and consult a doctor
It doesn't work after pressing
the '' '' button with batteries
installation
2726
Common question of blood pressure
measurement
Fault cause
Wrong
indication
Err 4
Err 5
Un ble to measure pressurea
Pressurizing error
Battery is running out
Replace new batteries
Battery is low
Replace new batteries
Err 8
Pressure exceeds the
maximum value (300mmHg)
Measure again please
Err 7
Cuff is too lose or fall off
Fasten cuff tightly
Err 6
Pressurizing error caused by
arm or body motion
Keep arm and body still
and measure again
Solution
Fasten cuff correctly
before measurement
Check if there is air
leakage from the cuff
The table below shows the possible fault displaying icon,
possible reason, and solutions.
Warning: If the situations cannot be solved or unexpected
problem happens, please consult the local distributor.