Axio Observer TECHNICAL DATA ZEISS
09/2017 431004-6044-008 9
English
Line voltage of external power supply unit of Axio Observer 7 materials ........... 100 V to 240 VAC ±10 %
A change of the line voltage is not required!
Mains frequency ................................................................................................................ 50 Hz 60 Hz
Power consumption Axio Observer 3, 3 materials and 5, 5 materials ..................................... max. 300 VA
Power consumption of external power supply unit of Axio Observer 7, 7 materials ............... max. 190 VA
Power supply unit for HBO 100
Operating area ................................................................................................................Enclosed rooms
Protection class ..................................................................................................................................... I
Ingress protection rating ................................................................................................................. IP 20
Line voltage ................................................................................................................ 100 V to 240 VAC
Mains frequency ................................................................................................................ 50 Hz 60 Hz
Power consumption if operating with HBO 100 ............................................................................ 155 VA
Fuses in accordance with IEC 127
Microscope stand Axio Observer 3, 3 materials and 5, 5 materials ................... T 5 A/H / 250 V, 5x20 mm
Power supply unit VP232-2 for Axio Observer 7, 7 materials .......................... T 4.0 A/H / 250 V, 5x20 mm
HBO 100 power supply unit (ballast unit) ...................................................... T 2.0 A/H / 250 V, 5x20 mm
Light sources
HBO 50W/AC mercury vapor short-arc lamp
Output ....................................................................................................................................... 50 W
Average service life ................................................................................................................... 100 h
HBO 103 W/2 mercury vapor short-arc lamp .................................................................................. 100 W
Desktop power supply unit for external filter wheel
Protection class ..................................................................................................................................... I
Desktop power supply unit input ........................................................ 100 V to 240 V ±10%, 50 60 Hz
Desktop power supply unit output ............................................................................ 24 VDC, max. 2.5 A
Optical-mechanical data
Stand with stage focusing ............................................ with coarse focusing drive approx. 2 mm/rotation
and fine focusing drive approx. 1/10 coarse/fine focus transmission ratio,
total travel approx. 10 mm, 13 mm also possible