AttentionUsers!!
FluoArcLamp:
TheFluoArclampmustbeonforatleast30minutesbeforebeingturned
offANDmustnotbeturnedonagainwithin1hourofbeingturnedoff.
1. Whenstarting,beforeyouturnontheFluoArclamp,makesureit
hasbeenOFFforatleast1hourfrom
previoususe.
2. Whenfinishing,beforeyouturnofftheFluoArclamp,makesureit
hasbeenONforatleast30minutes.
OilonObjectives:
Ifyouusedanoilobjective,pleaseremembertowipeoiloffwithdryLens
Paper.Donotuseethanol!
BookingTime:
Youareallowedtosignupforamaximumof3hoursperday.Please
makesureyouonlybooktimethatyouneed.Youareresponsiblefor
showingupforyourtimebooked.A$10.00feeappliesifyoudidnot
showupforbookedtime.
LogBook:
Allusersmustlogthetimespentonthestationaswellasthetimethe
fluorescencelampwasturnedon/offinthelogbook.
SaveYourData:
SaveallofyourworkinD:/drive,yourlabfolder,yourfolder.Files
foundontheDesktoporintheC:/drivewillbedeleteddaily.
Youareresponsibleforbackingupyourowndataatalltimes.Donot
relyonthiscomputerasstorageforyourdata.Onceyouhavebackedup
yourfileselsewhere,pleasedeleteyourfilesfromthiscomputer.