1) Before commencing treatment
- Adjust the slit lamp and eyepiece
- Main menu settings:
Applicator selection: Program selection:
VISULAS 532s Quick Instructions
2) Treatment STANDBY mode
Touch the button to switch from the main menu to STANDBY mode.
All laser parameters can now be changed again using the control knob by touching the corresponding parameter button
(power
also by pressing the control knob).
The pulse duration can be adjusted within the range 10 to 2500 ms or set to continuous ("cw"). For the "cw" setting, the
pulse duration must be rotated past the 2500 ms setting. "cw" will be displayed. The "cw" setting is only available for power
outputs of 500 mW or less.
The interval can, depending on the power and pulse duration, be varied between 100 and 6000 ms.
To change the aiming beam intensity , the aiming beam can be switched on in STANDBY mode by touching the
button.
The spot size can be adjusted on the slit lamp, depending on treatment and power, within the range 50 to 1000 µm.
These parameters can be configured in STANDBY or READY modes.
3) Touch the button to switch from STANDBY mode to READY mode.
The laser can only be fired using the foot switch in READY mode Æ TREATMENT!
Caution! Never fire the laser if the aiming beam is not clearly visible in the target area!
If, in READY mode, more than five minutes elapses without the laser being fired, the system will return automatically to
STANDBY mode.
4) A pulse counter is displayed. To reset the pulse counter to zero, touch the or button.
The slit lamp joystick control includes a micromanipulator, allowing quick and simple fine adjustment of the laser beam position (and the
slit lamp illumination). The focal plane is not changed, as moving the micromanipulator "towards" or "away from" the patient moves the
laser beam "upwards" or "downwards"!
000000-1404-671 VISULAS 532s Quick Instructions 29.05.2006
Display safety instructions concerning the physician's safety
filter for VISULINK, LIO and intraocular
p
robe a
pp
lications.
Button Function
Laser slit lamp
VISULINK
LIO
Intraocular probe
The P1, P2 and P3 programmes can be configured for each
applicator in program definition mode.
The power, pulse duration, interval and aiming beam intensity
laser settings are used each time the instrument is restarted.
Button Function
Factory set default program
P1, P2 and P3 are user-defined programs
Question and Answer IconNeed help?

Do you have a question about the Zeiss VISULAS 532s and is the answer not in the manual?

Zeiss VISULAS 532s Specifications

General IconGeneral
Laser TypeSolid-state laser
Wavelength532 nm
Aiming Beam Power< 1 mW
CoolingAir-cooled
Aiming BeamDiode laser
Electrical Requirements100 - 240 V, 50/60 Hz
ApplicationsRetinal photocoagulation
DisplayLCD
FootswitchYes
Treatment ModesSingle pulse

Summary

Commencing Treatment Setup

Applicator Selection

Selection of the correct applicator (slit lamp, VISULINK, LIO, intraocular probe) for treatment.

Program Selection

Choosing between factory default or user-defined programs (P1, P2, P3) for treatment.

Treatment STANDBY Mode Operations

Parameter Adjustment in STANDBY

Adjusting laser power, pulse duration, interval, and aiming beam intensity in STANDBY mode.

Spot Size Adjustment

Adjusting the laser spot size on the slit lamp, configurable in STANDBY or READY modes.

Transition to READY Mode and Firing

Switching to READY Mode

Transitioning from STANDBY to READY mode for laser firing via foot switch.

Safety Precautions for Firing

Crucial safety warning about firing the laser only when the aiming beam is visible.

Pulse Counter and Joystick Control

Pulse Counter Reset

Procedure to reset the pulse counter to zero using designated buttons.

Slit Lamp Joystick Functionality

Description of the micromanipulator for fine laser beam position adjustment.

System Navigation and Controls

Power-up and Main Menu Flow

Flow illustrating power-on, main menu, and access to diagnosis/system information.

Mode Switching Icons

Icons and descriptions for switching between system modes (Info, ECO, def).

Treatment Mode Icons

Icons and descriptions for switching between STANDBY and READY treatment modes.

Program Definition Mode

Functionality for generating user-specific programs.

Treatment Mode Displays

STANDBY Mode Display (LSL/VISULINK)

Interface display for STANDBY mode with parameters like pulse, interval, and power.

READY Mode Display (LSL/VISULINK)

Interface display for READY mode with parameters like pulse, interval, and power.

STANDBY Mode Display (LIO/ENDO)

Interface display for STANDBY mode with parameters like pulse, interval, and power.

READY Mode Display (LIO/ENDO)

Interface display for READY mode with parameters like pulse, interval, and power.