Make sure that during cleaning and disinfection no liquids penetrate the device. Do
not use sprays.
If during cleaning or disinfecting liquid penetrates the device, please put the unit out
of service, protect it from being used again and contact your service representative.
Make sure that when cleaning and disinfecting the labels of the device (such as
warnings, labels of control devices, identification plate) are not damaged.
Cleaning: In the event of visible contamination, the housing, the applicator and all
cables can be cleaned using commercially available alcohol-free plastic cleaners.
Wipe the surface until the dirt is removed, using a soft cloth soaked according to the
specifications of the manufacturer of the cleaning agent but not dripping wet.
Disinfection: We recommend that disinfection to be carried out at least once a
week, as well as in the event of evidence of possible contamination. Consult with
your health professional when doing so. Always perform cleaning prior to
disinfection.
Housing and applicator can be disinfected using disinfectant wipes. Use a
commercially available alcohol-free disinfectant for metal and plastic, with
bactericidal, veridical and fungicidal properties. Observe the application instructions
of the manufacturer. Wipe all surfaces using a cloth soaked according to the
specifications of the manufacturer of the disinfectant, but not dripping, or with cloth
pre-impregnated with disinfectant (wipes).
If applicable, also observe requirements for drying or post-cleaning.
Caution: If flammable solutions are used for cleaning and disinfection,
sufficient time must be allowed for the solutions to evaporate before using the
device. Otherwise, it may lead to inflammation!