Manufacturer´s EMC declaration
With regards to the EMC (electromagnetic compatibility), medical electrical devices such - as the
emFieldPro are subject to special safety measures and must be installed and commissioned in
accordance with the EMC instructions in the operating instructions or accompanying documents.
Portable and mobile HF communications equipment (e.g mobile telephones, cell phones) can impact
medical electrical devices.
The emFieldPro may only be operated using the original power cable indicated in the scope of delivery list.
Operating the device with another power cable can lead to increased emissions or reduced interference
resistance of the device.
Guidelines and manufacturer declaration - electromagnetic emitted interference
The emFieldPro device is intended for operation in an electromagnetic environment as indicated below. The
emFieldPro customer or user should ensure that it is operated in such an environment.
Electromagnetic environment - guideline
HF emissions in accordance with CISPR 11
The emFieldPro device uses HF energy
solely for its internal function. Therefore, its
HF emissions are very low and it is
improbable that it will interfere with
neighboring electronic devices.
HF emissions in accordance with CISPR 11
The emFieldPro device is suitable for use in
all facilities, including residential spaces and
areas directly connected to the public power
supply grid which also supplies buildings used
for residential purposes.
Harmonic emissions in accordance with
IEC 61000-3-2
Voltage fluctuations / flicker emissions in
accordance with IEC 61000-3-3
The device must not be used directly next to or stacked on/under other devices. If operation near or
stacked on/under other devices is necessary, the device should be observed to check for proper
operation in this setup.