Technical information
9
Page 13
Device type
Microwave treatment device
Operating frequency
2450MHz ± 25MHz
High
-
frequency power
at 60
unpulsed: max. 250W
pulsed: max. 125W pulse-pause ratio 1:1
pulsed low power: max. 50W pulse-pause ratio 1:2
Accuracy ±20%
The power figures relate to the average power output. The maximum power is
1200W.
Pulse frequency
0.5Hz, 1Hz
Accuracy ±5%
Treatment time
0 ... 30 minutes
Accuracy ±5 seconds
Mains connection
230V, 50Hz
Fuses
2 x T5 AL, 250 V
Power consumption
max. 900VA
Protection class
I (according to EN 60601
-
1)
Applied part
Type B (according to EN 60601
-
1), adjustable radiator
head
Environmental conditions
Temperature 10°C ... 30°C
Humidity 30% ... 75% without condensation
Air pressure 700hPa ... 1060hPa
Storage and transport
Temperature 0°C ... 50°C
Humidity 10% ... 90% without condensation
Air pressure 700hPa ... 1060hPa
only in original packaging
Dimensions
33cm x 42cm x 72cm (without radiator head and cruciform base)
47cm x 50cm x 83cm (without radiator head)
Weight
38.5
kg including radiator head
Disposal
The device or parts thereof must not be disposed of in
household waste.
Contact the relevant authorities or recycling firms regarding disposal.
Note:
Storage and transport only in original packaging.
Subject to technical changes.