Disinfection
10
Page 14
any maintenance and cleaning measures the device must always be
switched off at the main switch and the mains cable must be disconnected.
Make sure that during cleaning and disinfection no liquids penetrate the device. Do
not use sprays.
If during cleaning or disinfecting liquid penetrates the device, please put the device
out of service, protect it from being used again and contact your service
representative.
Make sure that when cleaning and disinfecting the labelling of the device (such as
warnings, labels of control devices, identification plate) is not damaged.
The device and its applied part are considered as uncritical in relation to hygiene
when used on non-injured and healthy skin.
radiator head
event of visible contamination, the housing and all cables can be cleaned
using commercially available alcohol-free plastic cleaners. Wipe the surface until the
dirt is removed, using a soft cloth soaked according to the specifications of the
manufacturer of the cleaning agent but not dripping wet.
We recommend that disinfection is to be carried out at least once a week, as well as
if there is any indication of contamination. Consult with your hygiene specialist when
doing so. Always perform cleaning prior to disinfection.
Housing and the adjustable radiator head can be disinfected using disinfectant
wipes. Use a commercially available alcohol-free disinfectant for metal and plastic,
with bactericidal, virucidal and fungicidal properties. Observe the instructions for use
of the respective manufacturer. Wipe all surfaces using a soft cloth soaked
according to the specifications of the manufacturer of the disinfectant, but not
dripping, or with cloths pre-impregnated with disinfectant (wipes).
Also observe requirements for drying or post-cleaning, where applicable.
Never spray disinfectant directly onto the device. Do not allow liquid to
penetrate the device or plug connectors!
Do not immerse the device or parts thereof in liquid!
Do not carry out steam, gas or gamma-ray sterilisation.
If flammable solutions are used for cleaning and disinfecting, sufficient time must be
allowed for the solution to evaporate before using the device. Otherwise, it may lead
to inflammation.
in a hygienic environment.