Problem Possible Cause Remedy
Calibration quality
of sensor pH <
25%
Impossible to
calibrate pH 4 (the
“Alarm/Alarme”
and “Hold/Pause”
diodes are
flashing)
Calibration quality
of the pH sensor
< 25%
The electrode
response is slow
problem with the sensor’s
porous element, dirt
deposits
the electrode is worn
buffer solution ineffective
problem with the
electrode bulb
the electrode is worn
static electricity on the
electrode
make sure that the electrode bulb is in good
condition, that it has not been allowed to dry
out of the water. As a last resort, clean by
soaking the electrode in water for a few hours
make sure the sensor’s porous element is in
good condition and wash the sensor with a
diluted acid-based solution
change the electrode
make sure the solution used is the pH 4 solution
check the pH of the buffer solution with an
electronic pH meter
take a new batch of pH 4 buffer solution and
begin calibration again
make sure that the electrode bulb is in good
condition, that it has not been allowed to dry
out of the water. As a last resort, clean by
soaking the electrode in water for a few hours
make sure the sensor’s porous element is in
good condition and wash the sensor with a
diluted acid-based solution
change the electrode DO NOT WIPE the
electrode with a cloth or paper but leave
to drain after rinsing
during the calibration phase DO NOT WIPE
the electrode with a cloth or paper but leave
to drain after rinsing