Installing the Sensor
Important: the length of pipe between the sensor and the injection point must not be less than 60cm.
The sensor must be vertical, or at 45° max and always above the pipe. In no circumstances should
it be positioned horizontally or below the pipe.
The sensor should be positioned in such a way as to analyze the pool water nearest to it and thus before
any treatment.
Installation Instructions
Correct Positions
Incorrect Positions