Warnings
9650-0912-06 Rev. G ZOLL R Series Operator’s Guide 1–17
To ensure patient safety, connect the R Series only to equipment with galvanically isolated
circuits.
Use only high-quality ECG electrodes. ECG electrodes are for rhythm acquisition only; you
cannot use ECG electrodes for defibrillation or pacing.
Do not use therapy or ECG electrodes if the gel is dried, separated, torn or split from the foil;
patient burns may result from using such electrodes. Poor adherence and/or air pockets under
therapy electrodes can cause arcing and skin burns.
Check the expiration date on the electrode packaging. Do not use electrodes after their
expiration date.
Excessive body hair or wet, diaphoretic skin can inhibit electrode coupling to the skin. Clip
excess hair and dry any moisture from the area where an electrode is to be attached.
Therapy electrodes should be replaced periodically during continuous pacing. Consult
electrode directions for proper replacement instructions.
Prolonged pacing (more than 30 minutes), particularly in neonates or adults with severely
restricted blood flow, may cause burns. Periodically inspect the skin under the electrodes.
Carefully route the patient cables to reduce the possibility of patient entanglement or
strangulation.
To avoid electrosurgery burns at monitoring sites, ensure proper connection of the
electrosurgery return circuit so that a return path cannot be made through monitoring electrodes
or probes.
During electrosurgery, observe the following guidelines to minimize electrosurgery unit (ESU)
interference and provide maximum operator and patient safety:
Keep all patient monitoring cables away from earth ground, ESU knives, and ESU
return wires.
Use electrosurgical grounding pads with the largest practical contact area.
Always ensure proper application of the electrosurgical return electrode to the patient.
Check electrical leakage levels before use. Leakage current may be excessive if more than one
monitor or other piece of equipment is connected to the patient.
Do not use the ZOLL OneStep Pacing cable (REF 1009-0913-02) or the ZOLL Multi-Function
Cable (REF 1009-0913-03) in a 220/240 VAC 60Hz power environment. Patient leakage
current may be excessive.

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB