Advisory Defibrillation Procedure
9650-0912-06 Rev. G ZOLL R Series Operator’s Guide 4–5
Units with the automatic charge option disabled will alternately display the messages
SHOCK ADVISED and PRESS CHARGE. Press the CHARGE button.
Regardless of the analysis result, the user can control the defibrillator manually. For example,
the user can defibrillate the patient even if the advisory function issues a NO SHOCK ADV.
message.
3 Press SHOCK
WARNING! Warn all persons in attendance of the patient to STAND CLEAR prior to defibrillator
discharge.
Do not touch the bed, patient, or any equipment connected to the patient during
defibrillation. A severe shock can result. Do not allow exposed portions of the patient’s
body to come in contact with metal objects, such as a bed frame, as unwanted pathways
for defibrillation current may result.
DEFIB 120J SEL.
Code
MarkerParam
Sync
On/Off
00:01
SHOCK ADVISED
ECG
Options Alarms
Report
Data

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB