Weight
Approximately 47g (with battery
installed)
Operating
temperature
5-45°C
Operating
relative
humidity
10-90% (non-condensing)
Hematocrit range 0-70%
Data port Micro USB
Limitations
The 2in1. MaX meter, test strips and control solution are been
designed, tested and proven to work together eectively to
provide accurate blood glucose measurements. Do not use
components from other brands.
• Fresh capillary blood may be collected into test tubes
containing sodium heparin, lithium heparin if the blood is
used within 10 minutes. Do not use sodium fluoride/oxalate
or other anticoagulants or preservatives.
• Use only with whole blood. Do not use with serum or plasma
samples.
• Very high (above 70%) hematocrit levels can cause false
results. Talk to your health care professional to find out your
hematocrit level.
• The system is tested to accurately read the measurement of
glucose in whole blood within the range of 10 to 600 mg/dL
(0.6-33.3 mmol/L).
• Fatty substances (triglycerides up to 3,000 mg/dL (166.7
mmol/L) or cholesterol up to 500 mg/dL (27.7 mmol/L))
have no major eect on blood glucose test results.
• The 2in1. MaX Blood Glucose Monitoring System has been
tested and shown to work properly up to 10,000ft (3,048
meters).