37
Interval memory mode
The weighing data is periodically stored in memory.
Required setting dout prt 3 Interval memory mode
dout data 2 Data memory function is used.
Stores weighing data.
dout int Interval time
Example “For periodical weighing without a computer command and to output
all of the data, to a computer, at one time”
10-6 Description of the Item “Data format”
A&D standard format 5if type 0
This format is used when the peripheral equipment can receive the A&D format. If an AD-8121B is
used, set the printer to MODE 1 or 2.
z This format consists of fifteen characters excluding the terminator.
z A header of two characters indicates the balance condition.
z The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus
sign is used.
z The unit, consisting of three characters, follows the data.
DP (Dump print) format 5if type 1
This format is used when the peripheral equipment can not receive the A&D format. If an AD-8121B
is used, set the printer to MODE 3.
z This format consists of sixteen characters excluding the terminator.
z A header of two characters indicates the balance condition. No overload header is used.
z The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is
not zero or overloaded.
z The unit, consisting of three characters, follows the data.
Unit
DataHeader
ST,+00012.7811g
C
R
L
F
Termi nator