46
Title block and end block
When a weight value is recorded as the GLP data, “Title block” and “End block” are inserted at the
beginning and at the end of a group of weight values, in the GLP report.
Note
To output the report to an AD-8121B , use MODE 3 of the AD-8121B.
Caution
If the data memory function is used, the “Title block” and “End block” can not be output.
Operation
1 With the weighing data displayed, press and hold the PRINT key until S5tart is displayed.
The “Title block” is output.
2 The weighing data is output according to the parameter setting of the data output mode.
3 Press and hold the PRINT key until recend is displayed. The “End block” is output.
When the setting is “info 1”: When the setting is “info 2”:
<TERM>
C
R
L
F
C
R
Space, ASCII 20h
Terminator, , or
Carriage return, ASCII 0Dh
Line feed, ASCII 0Ah
C
R
L
F
A & D<TERM>
MODEL GF-2000<TERM>
S/N 01234567<TERM>
ID ABCDEFG<TERM>
DA TE<TERM>
<TERM>
ST ART<TERM>
TI ME<TERM>
<TERM>
WT +1234.56 g<TERM>
WT +1234.61 g<TERM>
WT +1234.62 g<TERM>
WT +1234.63 g<TERM>
WT +1234.51 g<TERM>
WT +1234.53 g<TERM>
WT +1234.71 g<TERM>
WT +1234.64 g<TERM>
<TERM>
EN D<TERM>
TI ME<TERM>
<TERM>
SI GNAT URE<TERM>
<TERM>
<TERM>
----------------<TERM>
<TERM>
<TERM>
Weighing data
Time
Signature
End block
AD-8121 format
Manufacturer
Model
Serial number
ID number
Date
Time
Title block
A&D
MODEL
S/N
ID
DATE
START
TIME
GF-2000
01234567
ABCDEFG2 0 2 0
2008/04/18
11:12:09
WT +1234.56 g
WT +1234.61 g
WT +1234.62 g
WT +1234.63 g
WT +1234.51 g
WT +1234.53 g
WT +1234.71 g
WT +1234.64 g
END
03:50:300 PM
SIGNATURE
- - - - - - - -
General data format