Hardware Guide
Document Number: D1000092204 B 11
Figure 1-8 Gas panel
Source Exhaust System
The source exhaust system is a safety feature that isolates the sample vapors and
exhaust products from the laboratory environment. The ion source requires that
this exhaust system is properly connected and functioning. The exhaust pump
draws the solvent vapors from the enclosed source housing and delivers them to
the exhaust-waste-out fitting, and then to a waste bottle outside the instrument.
It is recommended that these vapors be passed through this waste bottle and
then vented to a fume hood or outside port.
WARNING! Hazardous Materials: Take all necessary precautions to ensure the safe
disposal of the source exhaust. Follow all applicable local regulations.
E
XH
A
U
S
T
S
U
PP
L
Y
M
IN
55
P
S
IG M
AX
6
0PS
IG
SHE
A
T
H
G
A
S
MA
X
105 P
SI
G
C
A
D
G
AS
MAX
60
PSI
G
C
A
D
G
AS
V
ALVE
WASTE OUT
E
XH
A
U
S
T
WAS
TE
OU
T
CURTA
IN
G
AS
S
UP
P
LY
Ma
x
6
0
P
S
IG
G
A
S
1
/
G
A
S
2
Max
1
0
5P
SIG
S
E
R
I
A
L
S
O
U
R
C
ES
A
U
XI/
O
I
E
E
E
-
4
88
207-
24
2V
50-60 H
z10
A
B
A
CK
I
N
G
P
UMP
Gas Panel
SHEATH
GAS
MAX 105 PSIG
CAD
GAS
MAX 60 PSIG
VALVE
WASTE OUT
EXHAUST
WASTE OUT
CURTAIN GAS
SUPPLY
MAX 60 PSIG
GAS 1 / GAS 2
MAX 105 PSIG
EXHAUST SUPPLY
MIN 55 PSIG MAX 60 PSIG