Hardware Guide
Document Number: D1000092204 B 15
Turbo V™ Ion Source Settings
Table 1-2 shows the recommended starting values for the X- and Y-axis
parameters. For more information, see the Turbo V™ Ion Source Operator’s
Guide.
Compound Parameters
In general, you can use the preset values for most of the parameters in Table 1-3.
For more information about the parameters, see the Analyst
®
software Help
system.
Table 1-1 Source/Gas Parameters
Parameter Value
Curtain Gas (CUR) 20
IonSpray Voltage (IS) 5000
Temperature (TEM) 700
Ion Source Gas (GS1) 60
Ion Source Gas 2 (GS2) 60
Table 1-2 Vertical and Horizontal Parameters
Parameter Value
X-axis 5
Y-axis 5
Table 1-3 Compound Parameters
Parameter Value
CAD Gas Use the preset value and optimize for your
compound.
DP (Declustering Potential) Use the preset value and optimize for your
compound.
EP (Entrance Potential) Use the preset value.
CXP (Collision Cell Exit
Potential)
Use the preset value and optimize for your
compound.
CE (Collision Energy) Use the preset value and optimize for your
compound.
IE1 (Ion Energy 1) — For
experienced instrument
operators only
Do not adjust.
IE3 (Ion Energy 3) — For
experienced instrument
operators only
Do not adjust.