32 AZ10 OXYGEN ANALYZER | COMBUSTION GAS ANALYSIS | OI/AZ10-EN REV. A
…12 Transmitter specification
Digital I/O
Number of digital I/Os
2
I/O configuration
User-configurable as either input or output
Input type
Volt-free contact
Output
Transistor switch capable of sinking 220 mA, low output,
< 2 V DC, switch voltage 30 V DC maximum
Isolation
Not isolated from each other or from
the microprocessor circuitry
Input functions
Automatic calibration start on falling edge

Automatic calibration start on rising edge

Automatic calibration stop on falling edge

Automatic calibration stop on rising edge








Output functions
Process alarm 1, 2, 3, 4
Calibration in progress
Calibration failed
Out of test gas 1
Out of test gas 2
Test gas 1 valve control
Test gas 2 valve control
Failure diagnostic
Out-of-specification diagnostic
Maintenance required diagnostic
Function check diagnostic
Hart communications
Version

Integration


Provide online/offline device configuration, online
monitoring of measurement values and diagnostic states
DTM
FDT v1.2.1 compliant
Works with FDT framework packages

EDD
Compliant with suitable framework tools

Languages
English
Calibration
Manual calibration



Automatic calibration


Calibration control
Front panel controls
Digital inputs
HART commands
User-defined schedule
Calibration scheduler
User-defined schedule enables automatic calibration
frequency to be set from 1 day to 12 months