18 AZ10 OXYGEN ANALYZER | COMBUSTION GAS ANALYSIS | OI/AZ10-EN REV. A
…7 Electrical connections
Cable preparation – non-AutoCal systems (both ends)/AZ10 probe to AZ10 transmitter
WARNING
Bodily injury.
• The 14-core AZ10 cable carries the screened signal wires
wires safely.
• Before preparing cable wires, refer to Table 1, page 19
to identify unused AutoCal wires that must be cut back
for non-AutoCal systems.
Referring to :
1 Before preparing cable ends, slide a cable gland onto each
end of the cable in the correct orientation.
2
3
exposed. Do not cut back Screen 2 drain wire.
4
Do not cut back Screen 1 drain wire.
5
6 Cut back all unused wires as instructed in Table 1, page 19
and cut the insulation back on each the remaining wires to
*Cut back
Figure 15 Cable preparation – non-AutoCal systems (both ends)/AZ10 probe to AZ10 transmitter
AZ10 probe
AZ10 transmitter
Screen 2
Screen 1 foil*
Screen 2 braid*
Screen 1
Violet
Grey
Red
Black
Green
White
Brown
Blue
Cut PVC outer cable sheath back
twisted together and
Cut back
White/Yellow
White/Black
White/Orange
White/Green
White/Red
White/Blue