EasyManuals Logo

ABB EasyLine EL3060 Series Instructions For Installation Start-Up And Operation

ABB EasyLine EL3060 Series
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #10 background imageLoading...
Page #10 background image
10 EasyLine EL3060 Series Commissioning Instructions
Information for the Installation, Maintenance and Repair of Electrical
Installations in Hazardous Areas
Installation to IEC/EN 60079-14 (VDE 0165 Part 1)
The electrical apparatus must be installed according to IEC/EN 60079-14
(VDE 0165 Part 1) "Electric Apparatuses for Potentially Explosive Gas
Atmospheres" Part 14: Electrical equipment in hazardous areas".
Equipotential Bonding
With respect to the equipotential bonding, the requirements of
IEC/EN 60079-14 and DIN VDE 0100 Part 410 "Protection against
electric shock" and part 540 "Earthing, PE conductors, equipotential
bonding conductors" must be complied with.
Electrostatic Charging
Electrostatic charging must be avoided. In this respect, the trade
association regulations for the "Avoidance of ignition hazards due to
electrostatic charges" (BGR 132) must be complied with.
Monitoring and Testing
Electrical equipment in hazardous areas must be monitored for its correct
condition. It must be checked by a qualified electrician as required, or at
least every three years, unless it is permanently monitored by a
responsible engineer.
Carrying out Work on Electrical Installations
Electrical installations in hazardous areas must be disconnected from the
power supply before any maintenance work is carried out on them. The
isolation point must be provided with a suitable warning sign e.g. "Do not
switch on - explosion hazard".
This does not apply to apparatus which may be opened under normal
operating conditions, e.g. recording instruments, or for which it is
expressly noted in the type examination certificate.
Carrying out Work on Intrinsically Safe Circuits
Work may also be carried out under voltage on intrinsically safe circuits in
hazardous areas.
However, the electrical data of the relevant test equipment (inductance,
capacitance, current value and voltage values) must be considered.
Particular attention is required when carrying out work on intrinsically safe
circuits, which have been installed in conjunction with zone 0.

Table of Contents

Other manuals for ABB EasyLine EL3060 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB EasyLine EL3060 Series and is the answer not in the manual?

ABB EasyLine EL3060 Series Specifications

General IconGeneral
BrandABB
ModelEasyLine EL3060 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals