13
1) Fit batteries to Organiser as described in Series 3 User guide page 3. Note
ERRATUM sheet, if any, packed with unit.
2) Use the
Esc
key to turn it on and then use and keys, followed by
Enter
to select the language required (most models
have this facility).
3) There may be a warning that the Backup battery is low, if so this can be ignored at this stage. (a new backup battery should
be tted later see manual). If this happens press the
Esc
key.
4) To ensure sufcient memory is available (128k Model), close down the applications
Data, Word, World, Calc and Agenda
as follows:
Move the highlight block with
or to application, press
Delete
, followed by
Y
.
Switch off by holding down and
1
together.
5) Assemble the Psion 3-link cable as described in the 3-link manual page 2 onwards and plug it into the connector on the left
side of the Series 3.
6) Press
Esc
to switch on again, this will give the system screen, showing the applications available. Press the Menu key, and use
the and keys to select the Apps menu (the centre menu). Select Install application by using the and keys then
press
Enter
.
Short cut: from the system screen, hold down and press
I
.
7) Press
key to select disk and then and to select C. The top line should now show Comms.app. If so press
Enter
.
(If not press the and then use the and keys to select it, then press
Enter
). A new icon will appear, a picture of the
organiser connected to a computer.
8) If necessary, use the cursor keys to highlight Comms[C] and press
Enter
to run the communications application. One of two
possibilities will occur:
a) The screen will blank briey and then a box containing a ashing cursor will appear, together with a message
Online…
Proceed to step 9.
b) A message
No system memory appears. See step 4 above.
Move the highlight back to
Comms[C] and repeat step 8.
9) Hold down and press
J
. Use the cursor ( and ) keys to make Serial Port settings as below:
Set Serial Port
* Baud rate 4800
* Data bits 8
* Stop bits 1
* Parity none
* Ignore parity yes
Press
Enter
when nished.
10)Hold down
and press
K
to set up handshakes as below.
Set Serial Handshake
* Xon/Xoff off
* Rts/Cts off
* Dsr/Dtr off
* Dcd off
11) Press the
and
O
keys, followed by , , to select drive Internal and then press .
Type in a suitable name, eg. MagMastr (8 characters max.) and press
Enter
, to save the settings for future use.
The Series 3 is now ready for use with the MagMaster transmitter, as described in this Conguration Manual.