EasyManua.ls Logo

ABB TTH300 Series - HART Output Characteristics; PROFIBUS PA Output Signals

ABB TTH300 Series
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TTH300 HEAD-MOUNT TEMPERATURE TRANSMITTER | OI/TTH300-EN REV. E 25
Output – HART®
Note
The HART® protocol is an unsecured protocol, as such the
intended application should be assessed to ensure that these
protocols are suitable before implementation.
Transmission characteristics
• Temperature linear
• Resistance linear
• Voltage linear
Output signal
Configurable 4 to 20 mA (standard)
Configurable 20 to 4 mA
(Dynamic range: 3.8 to 20.5 mA in accordance with NE 43)
Simulation mode
3.5 to 23.6 mA
Induced current consumption
< 3.5 mA
Maximum output current
23.6 mA
Configurable error current signal
Overrange 22 mA (20.0 to 23.6 mA)
Underrange 3.6 mA (3.5 to 4.0 mA)
Output – PROFIBUS PA®
Note
The PROFIBUS PA® protocol is an unsecured protocol, as such
the intended application should be assessed to ensure that
these protocols are suitable before implementation.
Output signal
PROFIBUS – MBP (IEC 61158-2)
Baud rate 31.25 kBit/s
• PA-Profile 3.01
FISCO compliant (IEC 60079-27)
• ID-Number: 0x3470 [0x9700]
Error current signal
FDE (Fault Disconnection Electronic)
Block structure
• Physical Block
Transducer Block 1 – Temperature
Transducer Block 2 – HMI (LCD indicator)
Transducer Block 3 – enhanced diagnosis
Analog Input 1 – Primary Value (Calculated Value*
)
Analog Input 2 – SECONDARY VALUE_1 (Sensor 1)
Analog Input 3 – SECONDARY VALUE_2 (Sensor 2)
Analog Input 4 – SECONDARY VALUE_3 (reference junction
temperature)
Analog Output – optional HMI display
(Transducer Block 2)
Discrete Input 1 – extended diagnosis 1
(Transducer Block 3)
Discrete Input 2 – extended diagnosis 2
(Transducer Block 3)
* Sensor 1, Sensor 2 or difference or mean

Table of Contents

Related product manuals