EasyManua.ls Logo

ABB TTH300 Series - Page 42

ABB TTH300 Series
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
42 TTH300 HEAD-MOUNT TEMPERATURE TRANSMITTER | OI/TTH300-EN REV. E
… 11 Operation
… Parameter description HART®
Menu / parameter Description
R-Connection Sensor connection type relevant for all Pt, Ni, Cu resistance thermometers
two-wire: Sensor connection type in two-wire technology
three-wire: Sensor connection type in three-wire technology
four-wire: Sensor connection type in four-wire technology
2-wire Resistance Sensor line resistance relevant for all Pt, Ni, Cu resistance thermometers with a two-wire circuit
Value range: 0 to 100
Thermocouple RJ Internal: Use of internal reference junction of the transmitter when using thermal compensating cable.
External - fixed: Use of external fixed reference junction of the transmitter when constant thermostat temperature is
used (can be set with ext. Thermocouple RJ).
Without: no Thermocouple RJ
Sensor 1: Use of sensor 1 as reference junction for sensor 2
ext. Thermocouple RJ Relevant for external reference junction, specification of constant external reference junction temperature
Value range: -50 to 100 °C
… / Device Setup / Measured range
Engineering Unit Selects the physical unit for the sensor measuring signal
Units: °C, °F, °R, K, user, mV, Ω, mA
Measured Range Begin Defines the value for 4 mA (adjustable)
Measured Range End Defines the value for 20 mA (adjustable)

Table of Contents

Related product manuals