EasyManua.ls Logo

ABB WaterMaster FET100 - Terminal Connections and Cable Prep

ABB WaterMaster FET100
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WaterMaster FET100
Electromagnetic floweter | Transmitter 3 Electrical Installation
10 OI/FET100–EN Rev. A
3.2 Transmitter Terminal Connections
Referring to Fig. 3.3:
1. Slacken (but do not remove) the four transmitter cover screws
A.
2. Remove the transmitter cover.
3. Check that the power indicator LED
B on the backplane is not lit.
4. If screws
C are not visible, access them by gently pulling the rotation lock D back and rotating the
cartridge
E until the cartridge screw access holes align with the cartridge screw heads.
5. Slacken the three cartridge screws and lift the cartridge
F away from the housing.
Warning. Isolate the transmitter from power supplies before removing the cover.
Fig. 3.3 Accessing the Transmitter Terminals
Warning. If the power indicator LED B is lit, the transmitter is still powered up. Before
continuing, isolate the transmitter power supply.
A
C
C
A
A
A
C
E
D
F
B