WaterMaster FET100
Electromagnetic floweter | Transmitter 3 Electrical Installation
OI/FET100–EN Rev. A 13
3.5 Output Connections
3.5.1 Frequency Outputs
Caution.
Inductive loads must be suppressed or clamped to limit voltage swings.
Operation of outputs is programmable.
External isolators are not normally required as the pulse and alarm circuit is electrically separated
from all other WaterMaster connections.
Fig. 3.6 Frequency Output Connections
Note. Outputs 1 and 2 are polarity sensitive. The common (negative) connection for these outputs is
designated 'COM'.
XX XX 42
52
62
41 51 61XX XX
PLC or Datalogger
COM
O/P1 O/P2
Terminal connection IDs are
HART- / PROFIBUS-variant
dependent