199
Technical data
13
Performance evaluation
The performance data for the Accu-Chek
Compact Plus system (Accu-Chek
Compact Plus meter together with Accu-Chek
Compact test strips) were determined using
capillary blood from patients with diabetes
(method comparison, accuracy), venous blood
(repeatability) and control solutions
(reproducibility). The system is calibrated with
venous blood containing various levels of
glucose. The reference values are obtained
using the hexokinase method. The method
comparison was conducted by comparing the
results from the system with those of the
hexokinase method with deproteination using
an automatic analyzer. The reference method
is traceable to a NIST standard using the
ID-GCMS method, which is the method of the
highest metrological quality.
The Accu-Chek Compact Plus system complies with the requirements of
EN ISO 15197.
man_06919570001_01_EnCan.indd 199 18.01.2013 18:39:07